Motor settings and file selection

Check speed, torque, and the correct file sequence before the first instrument enters the canal.

Chairside sequence

Step by step

  1. 01

    Identify the instrument

    Confirm the system, size, taper, and intended role. Do not select a preset by a similar product name.

  2. 02

    Verify the instructions

    Check the current manufacturer instructions for speed, torque, motion, and single-use requirements.

  3. 03

    Establish the glide path

    Motor settings cannot compensate for an unconfirmed path or coronal interference.

  4. 04

    Monitor each engagement

    Use light pressure, clean the flutes, and stop if the file binds or feedback changes.

Use the current instructions supplied with the exact instrument. Settings are not interchangeable between file systems.

Sequence follows anatomy

The NZ ET example progresses from Micro-Path 13/.03 to ET 20/.04 and ET 25/.04. The PT guide describes SX, S1, and S2 followed by an appropriate finishing file. These are source sequences, not a requirement to take every instrument through every canal.

Inspection and use

Inspect files before and during treatment. Visible distortion is a reason to discard an instrument, but fatigue damage may not be visible. Follow the instrument’s labeling for use and disposal; do not treat the NZ source’s general inspection advice as permission to reuse a single-use USA file.

Reference table

Settings listed on the EndoTech NZ source page — reference only; confirm the USA instrument instructions before use.
NZ source systemSpeedTorqueRole
Micro-Path300 rpm2.0–2.5 N·cmGlide path
ET350 rpm3.0 N·cmConstant-taper shaping
PT300 rpm3.0 N·cmProgressive-taper shaping

Sources & further reading

Read the source material

Adapted for the EndoTech USA education library. NZ reference documents retain their original edition and branding.