Education/ET constant-taper protocol
TransformSTM ET
Establish the glide path, then select .04 or .06 taper. Follow the illustrated ET route below and choose the preparation that fits the canal.
Compatible with X7 and HyFlex
Third-party names identify file systems and workflow compatibility only. No affiliation, sponsorship or endorsement is implied. Trademark notice.
Illustrated sequence / protocol card page 3
Glide → open → prepare
15/.03 → 20/.06 → 17/.04 → 25/.04
-

MicroPath 15/.03
Glide path
To length -

ET 20/.06
Open / preflare
To midroot -

ET 17/.04
Initial preparation
To length -

ET 25/.04
Final preparation
To length
Keep preflare controlled. Use 20/.06 toward mid-root only within an established, sufficiently wide straight coronal path. Stop before curvature or resistance; mid-root is not a mandatory depth.
Sequence + clinical decisions
Step by step
Choose the route that fits the canal.
Assess anatomy, isolate with a rubber dam, scout with a small hand file and establish working length (WL).
- 01
Glide path
After hand scouting, select MicroPath 15/.03 or 17/.03 for routine glide preparation; use 13/.03 for complex or narrow paths. Confirm a smooth, reproducible path to working length.
- 02
Open the entrance
Use the dedicated 17/.08, 17 mm opener only in the established coronal path. Brush away from the furcation and stop before curvature or resistance. The illustrated 20/.06 preflare is an optional route for a confirmed, sufficiently wide straight coronal path.
- 03
Initial preparation · 17/.04
Follow the glide path with light pressure, withdrawing frequently to clean and inspect. Reconfirm working length after coronal enlargement.
- 04
Develop the shape · 25/.04
Try 25/.04 passively to working length. If it cannot advance, use 20/.04, then return to 25/.04. Never drive through resistance.
- 05
Choose the endpoint
Select further enlargement from the apical assessment and anatomy. Use 25/.06 or larger .04/.06 files only when justified.
- 06
Complete treatment
Record final size, taper and working length. Complete disinfection, check the selected cone at working length, then obturate under length control.
Optional 20/.06 preflare
A size 20/.06 file taken toward mid-root before 17/.04 and 25/.04 can improve cone fit in the final preparation. Use it only within the confirmed straight coronal path and verify the final cone at WL.
Alternative crown down
ET files can also be used in a crown-down technique: 25 → 20 → 17, in .04 or .06 taper.
A dedicated instrument
Orifice opener
The orifice opener is a unique 17/.08, 17 mm file. It was created for clinicians who find larger-tipped orifice shapers difficult to enter into smaller calcified orifices, helping position the instrument when redefining a difficult canal entrance.
Use it only in the established coronal path. Brush away from the furcation and stop before curvature or resistance.

Motor setup
EndoTech settings
Rotary motion · torque in N·cm
| Instrument | Speed | Torque |
|---|---|---|
| ET shaping files | 400–500 rpm | 2 N·cm |
| Orifice opener · 17/.08 · 17 mm | 400–500 rpm | 2 N·cm |
| MicroPath 15/.03 and 17/.03 | 400–500 rpm | 2 N·cm |
| MicroPath 13/.03 | 400–500 rpm | 1 N·cm |
Read the instrument designation on the packaging before use. ET shaping files are available in sizes 15 to 45. Use EndoTech settings and the current instructions for the instrument in use.
Inspect between instruments
Keep the pathway reproducible
- Withdraw, clean and inspect the flutes between instruments.
- Irrigate, recapitulate with a small hand file and re-irrigate.
- Confirm a smooth return to working length. Rebuild the path if it is no longer reproducible.
- Record final size, taper and working length; verify the final cone at working length before obturation.
Keep the sequence to hand
ET protocol card
The three-page card includes the clinical sequence, technique notes and illustrated four-file pathway.
Adapted from the approved EndoTech card, Edition 09, 8 October 2026. The PDF retains its NZ branding. Instrument illustrations are not a dimensional scale.